A hospital needed to confirm that a nearby railway and MRI suite would not stop a new radiotherapy linac meeting its manufacturer’s static magnetic field limit.

The challenge. A hospital planned a new bunker for a radiotherapy linear accelerator (linac). The manufacturer requires the machine to sit outside any static magnetic field above 100 µT. The site lies close to a railway line and an existing MRI suite, both potential sources of strong static fields. The architects needed to know whether the chosen location would work before construction started.

What we did. We logged the static magnetic field at twelve positions with a three-axis magnetometer, including the bunker corners, the ground nearest the railway and a transect towards the MRI room. We established the local background and assessed the effect of passing trains on the Earth’s field.

Radiotherapy linear accelerator treatment room
A linac must be sited away from strong static magnetic fields.

The outcome. Fields across the bunker footprint were 38–53 µT, well under the 100 µT criterion, giving a margin of at least 47 µT at the isocentre. The only reading above 100 µT was at the MRI room wall, outside the footprint. The location was confirmed as suitable, with follow-up checks recommended once the structural steel is in place and before commissioning.