Buried power cables were disturbing electron microscopes on a production floor. We measured how far shielding could realistically bring the fields down.
The challenge. A high-technology manufacturer was seeing interference on electron microscopes on its production floor. The suspected cause was 50 Hz magnetic fields from buried 400 V circuits beneath the floor, at around 1 µT at floor level. The company wanted to know whether those fields could be brought down towards 0.1 µT.
What we did. We measured the fields over the buried conductors and reported on the options for reducing them. We then assessed the shielding options and measured their real-world performance on site, testing mu-metal and several thicknesses of a commercial shielding fabric over the problem conductors.

The outcome. The measurements showed that a tenfold reduction was not practical at floor level. A threefold reduction was achievable, and we set out what that would involve in cost and disruption, so the client could weigh it against other options. Each stage was reported with a short slide summary for the decision-makers.